DuaLogR Thermocouple Thermometer


DiGi-Sense® DuaLogR® Thermocouple Thermometer and Datalogger (60010-50)

  • Log up to 1000 readings
  • Transfer data to printer or computer by infrared output
  • Accepts up to two J, K, T, E, R, S, N or B thermocouples
  • Accuracy: ±0.7°F (±0.4°C) above -238.0°F (-150.0°C) plus ±0.1% of reading
  • Intrinsically Safe
  • Bright, easy-to-read LCD
  • Rated IP 54 for splash & dust resistance
  • ABS/polycarbonate impact resistant case
  • Fold out stand for bench top operation
  • UL approved for Class I, Division 1; Groups A, B, C, and D hazardous (classified) locations
  • Meets CE requirements for electromagnetic compatibility

DiGi-Sense® DuaLogR thermometer features real time data logging and infrared data output capabilities. Meter accepts up to two thermocouple inputs. Choose from J, K, T, E, R, S, N or B thermocouples. Display temperatures in °F, °C, °R or K scales. Resolution and input filtering are selectable to improve readability during rapidly changing temperatures. Display minimum, maximum and T1/T2 differential readings simultaneously using one or two probes on selectable, easy-to-read, multi-data LCD.

Manually store or automatically log up to 1000 sets of readings in real time. Log interval is selectable from 1 second to 60 minutes. Order the optional infrared printer or the RS-232 interface adapter to download readings directly to your computer. Besides being battery powered, meter can be connected by an optional AC adapter for continuous operation.

 

Two point field calibration feature for offset and slope. Allows user to calibrate the probe to one or two known temperatures. Meter automatically stores and adjusts for inherent probe error enabling unparalleled system accuracy. Calibration lock out feature protects NIST traceable factory settings and prevents tampering with field calibration values.

Meter Specifications

Range J

-328 to 1832°F (-200 to 1000°C)
Range K -418 to 2501°F (-250 to 1372°C)
Range T -418 to 752°F (-250 to 400°C)
Range E -418 to 1832°F (-250 to 1000°C)

Range R

32 to 3214°F (0 to 1768°C)
Range S 32 to 3214°F (0 to 1768°C)
Range N -418 to 2372°F (-250 to 1300°C)
Range B 392 to 3272°F (200 to 1800°C)
Accuracy Type J, K, T, E, and N:
Above— 238°F (-150°C) ± 0.7°F (±0.4°C)
  plus 0.1% of reading
Below— 238°F (-150°C) ± 2°F (±1°C)
  plus 0.25% of reading
  Type R, S, and B
± 2°F (±1°C) ±0.1% of reading
Resolution 0.1° between -150° and 999.9°
1.0° below -150 and above 999.9°
Display Multi-line Three 4-digit custom custom LCD