DuaLogR
Thermocouple Thermometer
DiGi-Sense®
DuaLogR® Thermocouple Thermometer and Datalogger (60010-50)
Log
up to 1000 readings
- Transfer data to printer or computer by
infrared output
- Accepts up to two J, K, T, E, R, S, N
or B thermocouples
- Accuracy: ±0.7°F (±0.4°C) above -238.0°F
(-150.0°C) plus ±0.1% of reading
- Intrinsically Safe
- Bright, easy-to-read LCD
- Rated IP 54 for splash & dust resistance
- ABS/polycarbonate impact resistant case
- Fold out stand for bench top operation
- UL approved for Class I, Division 1; Groups
A, B, C, and D hazardous (classified) locations
- Meets CE requirements for electromagnetic
compatibility
DiGi-Sense®
DuaLogR
thermometer features real time data logging and infrared data output capabilities.
Meter accepts up to two thermocouple inputs. Choose from J, K, T, E, R,
S, N or B thermocouples. Display temperatures in °F, °C, °R or K scales.
Resolution and input filtering are selectable to improve readability during
rapidly changing temperatures. Display minimum, maximum and T1/T2 differential
readings simultaneously using one or two probes on selectable, easy-to-read,
multi-data LCD.
Manually
store or automatically log up to 1000 sets of readings in real time. Log
interval is selectable from 1 second to 60 minutes. Order the optional
infrared printer or the RS-232 interface adapter to download readings
directly to your computer. Besides being battery powered, meter can be
connected by an optional AC adapter for continuous operation.
Two point field calibration feature for offset and slope. Allows user
to calibrate the probe to one or two known temperatures. Meter automatically
stores and adjusts for inherent probe error enabling unparalleled system
accuracy. Calibration lock out feature protects NIST traceable factory
settings and prevents tampering with field calibration values.
Meter Specifications
| Range
J |
-328
to 1832°F (-200 to 1000°C) |
| Range
K |
-418
to 2501°F (-250 to 1372°C) |
| Range
T |
-418
to 752°F (-250 to 400°C) |
| Range E |
-418
to 1832°F (-250 to 1000°C) |
| Range
R |
32
to 3214°F (0 to 1768°C) |
| Range
S |
32
to 3214°F (0 to 1768°C) |
| Range
N |
-418
to 2372°F (-250 to 1300°C) |
| Range B |
392
to 3272°F (200 to 1800°C) |
| Accuracy |
Type
J, K, T, E, and N:
Above— 238°F (-150°C) ± 0.7°F (±0.4°C)
plus 0.1% of reading
Below— 238°F (-150°C) ± 2°F (±1°C)
plus 0.25% of reading
Type R, S, and B:
± 2°F (±1°C) ±0.1% of reading |
| Resolution |
0.1°
between -150° and 999.9°
1.0° below -150 and above 999.9° |
| Display |
Multi-line
Three 4-digit custom custom LCD |
|